  • The Network Mode is used to switch between EIP or ModBus TCP network.

  • EIP Setup
    Input Assembly Instance = 0x64 (100), size = 20 bytes
    Output Assembly Instance = 0x70 (112), size = 20 bytes
    Configuration Assembly Instance = 0x96 (150), size = 44 bytes.
    Network command interface
    The PLC to Hi6200 "network command interface" uses 20 bytes (10 MODBUS registers) in and 20 bytes out.
    • Output
      Bytes 0,1(MODBUS registers 0): command register
      Bytes 2,3 (MODBUS register 1): aux command information register
      Bytes 4,5,6,7 (MODBUS registers 2,3): parameter value
      Bytes 8,9 (registers 4): Parameter number. See the diagnostics/parameters web page for a list of parameter numbers.
    • Input
      Bytes 0,1(MODBUS registers 0): command register
      Bytes 2,3 (MODBUS register 1): Sample Counter and command status register
      Bytes 4,5,6,7 (MODBUS registers 2,3): parameter value
      Bytes 8,9 (registers 4): Parameter number. See the diagnostics/parameters web page for a list of parameter numbers.
      Bytes 10,11 (registers 5): instrument status register
      Bytes 12,13,14,15 (registers 6,7): net weight register
      Bytes 16,17,18,19 (registers 8,9): gross weight register
    The master sends a 'command' by writing a value the command register. Here is a list of command numbers:
    • 0x00 : READPARAMCMD. To read a parameter, write 0 to the command register (register #0), and write the parameter number in the parameter number registers 4. The parameter value may then be read from registers 2 and 3. This value may be in integer or floating point format, depending on the parameter. The instrument status register in the reply will contain the system status word.
      • Status word bit 0: A/D error.
      • Status word bit 2(0x04): Motion status.
      • Status word bit 7(0x80): Not Found - the requested parameter number does not exist.
    • 0x01 : ZEROCMD. Write 1 to the command register to ZERO the gross weight. The status register will read 0 if this command succeeds. Error codes are 1 (A/D error), 3 (out of tolerance) and 4 (motion).
    • 0x02 : TARECMD. Write 2 to the command register to Tare the net weight. The status register will read 0 if this command succeeds.Error codes are 1 (A/D error) and 4 (motion).
    • 0x64 : CALLOWCMD. Perform the low step of a calibration.
    • 0x65 : CALHIGHCMD. Perform the high step of a calibration.
    • 0x66 : C2CALCMD. Do a C2 calibration.
    • 0x92 : WRITEINTEGERCMD. Set the value of an integer parameter. Write the parameter number in the parameter number registers and the desired value in the parameter value registers.
    • 0x93 : WRITEFLOATCMD. Set the value of a floating point parameter. Write the parameter number in the parameter number registers and the desired value in the parameter value registers.
    • 0x96 : WRITENONVOLCMD. Write 0x96 to the command register to save parameters in non-volatile memory.