SB
 
  • The Loadcell Sensitivity parameter should be set greater than or equal to the sensitivity of the load cell you are using.
  • A C2 Calibration can be done if you are using Hardy C2 load sensors. The Num Devices shows how many C2 sensors are connected.
  • The Ref Weight is the test weight used during the low step of calibration
  • C2 load sensors are programmed to produce correct weight at STANDARD GRAVITY. You may use the Gravity Correction parameter to adjust to your local gravity. This gravity correction parameter may be between 0.900000 and 1.100000
  • A Traditional Calibration is done with test weights. There are 2 steps: a "Cal Low" step done with the Ref Weight on the scale, and a "Cal High" step.
  • The Span Weight is the weight on the scale at the high step of calibration. The low step should be done prior to the high step.
  • The Multi-Point Calibration is supported in Hi6200. Please click the "Multi-Point Calibration" link to do the it for nonelinear load cells.

Home - Configuration - Monitor - Wizard - HI 6200 Series Online Support Site

Hardy Process Solutions
San Diego, California, USA
Tel. +1 (858) 278-2900 ▪ (800) 821-5831
Fax +1 (858) 278-6700
www.hardysolutions.comhardyinfo@hardysolutions.com
hardysupport@hardysolutions.com

homeconfigurationMonitorWizardhardy